Ad ML-50 Manuel d'utilisateur Page 39

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 74
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 38
MS-70 / MX-50 / MF-50 / ML-50 37
7.2.3. Step Drying
Standard Mode, Automatic Mode or Manual Mode
Refer to page 39 for Timer Mode
This explanation uses the following parameters and sample displays of MX-50.
"Temperature 1" can be set higher than "temperature 2" in step drying.
Common Items
Program number ..........................................5 ( PROG 5 )
Drying program Heating pattern....................Step drying ( )
Drying temperature 1 ..........120°C
Drying temperature 2 ..........160°C
Time.....................................5.0 minutes
Measurement Unit..............................................Moisture content
Minimum scale value during measurement ..0.01 %
Minimum scale value of gram display...........0.001 g
Sample quantity............................................Approximately 5 g
Data memory function ..................................Not used
Items for Standard Mode
Analysis mode ..............................................Standard mode (Symbol: 5td )
ACCURACY .................................................MID.
Analyzing mode to finish measurement............Automatic setting by ACCURACY
Items for Automatic Mode
Analysis mode ..............................................Automatic mode (Symbol: U-a )
Analyzing mode to finish measurement............0.05 %/min
Items for Manual Mode
Analysis mode ..............................................Manual mode (Symbol: U-m )
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
2. Press the PROGRAM key and press the or
key to select a program number.
3. Press the ENTER key to use the number.
4. The analyzer displays end and returns to the
weighing mode.
Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
% MOIST
D /W/D
PROG 1
0.000g
PROG "5
end
PROG 5
0.000g
Program number
Time
Drying temperature 2
Drying temperature 1
Stage 1 Stage 2
(ML-50 can not select step drying)
Vue de la page 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 73 74

Commentaires sur ces manuels

Pas de commentaire