Calibration Report Example for the Weighing Sensor Adapted to GLP, GMP and ISO
Preset the following parameters to print data to AD-8121B
Device Parameter Description
Interval pU5e 1 1 Approx. two seconds interval in each line.
Analyzer
Output format
adapted to GLP,
GMP and ISO
info 1
or
info 2
Calibration report is printed at "Calibration of the
Weighting Sensor" and "Calibration of Drying
Temperature".
AD-8121B
MODE 3 Dump print (Received data is printed as it is)
1: Factory settings
...........A & D .........
Manufacture
MODEL......MX-50 .........Model
S/N.....K1234567 .........Serial number
ID.......LAB-123 .........ID number
DATE..2004/09/30 .........Date
TIME....13:57:24 .........Time
CALIBRATED
..........WEIGHT
CAL.WEIGHT
.......20.000..g
SIGNATURE
.........Signature
- - - - - - - -
} Calibration type
}
Calibration mass