MS-70 / MX-50 / MF-50 / ML-50 27
7.1.1. ACCURACY of the Standard Mode and Quick Mode
ACCURACY of measurement can be set either HI , MID. or LO .
The sample quantity is automatically selected by ACCURACY.
The termination value of the analyzing mode is automatically selected by ACCURACY
and minimum scale value of % display.
The analyzing mode is the program to finish the measurement when a change of
moisture content per one minute becomes smaller than a preset termination value.
The settings are as follows: Specify an ACCURACY.
Drying rate: Change of moisture content per one minute [%/min]
Standard Mode
ACCURACY
Model Minimum scale HI MID. LO
0.001 % 0.01 %/min 0.02 %/min 0.05 %/min
0.01 % 0.02 %/min 0.05 %/min 0.10 %/min
MS-70
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.01 % 0.02 %/min 0.05 %/min 0.10 %/min
MX-50
0.1 % 0.10 %/min 0.20 %/min
0.05 % 0.05 %/min 0.10 %/min
0.1 %
MF-50
1 %
0.1 %
Preset
Termination
value
ML-50
1 %
0.10 %/min 0.20 %/min
0.50 %/min
Sample quantity 10 g 5 g 1 g
Use Precise result
Quick measurement
Quick Mode
ACCURACY
Model Minimum scale HI MID. LO
0.001 % 0.02 %/min 0.05 %/min 0.05 %/min
0.01 % 0.05 %/min 0.10 %/min 0.20 %/min
MS-70
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.01 % 0.05 %/min 0.10 %/min 0.20 %/min
MX-50
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.05 %
0.1 %
MF-50
1 %
0.10 %/min 0.20 %/min 0.50 %/min
0.1 %
Preset
Termination
value
ML-50
1 %
0.20 %/min 0.50 %/min 1.00 %/min
Sample quantity 5 g 2 g 1 g
Use Precise result
Quick measurement